SIRONA UNIVERSAL DAC CLEANING SYSTEM

At Unique Technology we provide onsite repair / service + Calibration and validation for Sirona DAC Universal autoclave 

Repair DAC Universal

Many dental practice have an issues with there DAC universal autoclave and looking for expert who can solve those issues. At Unique Technology, We offer onsite repair DAC universal autoclave and training of the staffs with high efficiently and low cost

What the DAC Universal do?

Using the DAC Universal reduces the time spent cleaning and lubricating the handpieces, contra angles and turbines manually. Also, as the DAC Universal is a highly efficient autoclave, it ensures a fast process, thereby reducing the time spent on cleaning, maintenance and sterilisation procedures in the clinic. The assistant will spend less time in the sterilisation room and more with the patients, helping the doctor.

What are the different types of servicing

There are two types of servicing

  1. Annual service
  2. Major service

Contact us for more info

DAC Universal training

We always can provide on-site training for the staff, you can book us for training the staff or after finishing the service.

training will include but not excluded

  1. How to use the autoclave safely
  2. how to maintain the autoclave daily, weekly, monthly
  3. what are the common errors and how to solve them

Errors list and solutions

ErrorDescription
64The safety system has been activated -> Press C -> Restart the autoclave
65The safety system is defective -> Please contact us
66The temperature during sterilization is too high -> Please contact us Contact Us
67The time of sterilization has been less than 3 minutes -> The instruments should
be sterilised again -> Please contact us if this error appears more than
once.
68The pressure is too high during pressure test  Please Contact Us
70The pressure level does not decrease fast enough after pressure test -> Empty
the chamber for water -> Clean or change the filter -> Check that the drain tube is
not blocked or broken (note that the drain tube must be of heat resistant material,
e.g. PTFE) -> Empty the drain tank if full -> Please Contact Us
71The autoclave was blocked due to use of bad water -> Control the level of microsiemens in the water  Empty the watertank -> Fill in fresh water (below 3MS)
72The closing system was blocked on the way up -> Remove possible blocking for
the lift -> Please Contact Us
73The sterilization temperature has not been reached fast enough before the first
back-flush -> Pour in water in case of low level -> Please Contact Us
74The right temperature for building up pressure for sterilization has not been
reached fast enough -> Pour in water if low level -> If the autoclave is placed in a
cold spot remove it to surroundings at room temperature -> Press C to reset and
then restart -> Please Contact Us
75The right sterilization temperature has not been reached fast enough -> Pour in
water if low level -> If the autoclave is placed in a cold spot move it to a
surrounding at room temperature -> Press C to reset and then restart -> Please
Contact Us
76Valve error -> Please Contact Us or try with only 5 instruments on the lid
77The autoclave does not lubricate -> Please Contact Us
78Valve error -> Please Contact Us or try with only 5 instruments on the lid.
79Valve error -> Please Contact Us or try with only 5 instruments on the lid.
81Low level in either the water tank or the oil bottle  Pour water into the water tank
or change the oil bottle -> Please Contact Us
82The temperature in the steam generator is too high -> Please Contact Us
83Temperature in chamber is too high -> Please Contact Us
84The sterilization temperature has not been maintained during sterilization -> Pour
in water, if needed -> Please Contact Us
  • What is the main function of DAC Universal?

    The DAC UNIVERSAL automatically cleans, lubricates and sterilises turbines, handpieces and contra-angles (in the
    following designated instruments).
    In approx. 12 minutes 6 instruments will be cleaned, lubricated and sterilised and are ready for use again after cooling
    down.

  • what is the quality of the water we should use in the DAC Universal?

    It is necessary to use distilled or demineralised water of max. 3 MS/cm (Micro Siemens).

  • How to Increasing oil expel?

    You can increase the oil expel in the DAC Universal by fallow the fallowing steps
    1-Enter the menu system by touching the “Menu/Enter” button.
    2-Press the “Up” or the “Down” button until you reach “Settings”.
    3-Press the “Menu/Enter” button to select “Settings” mode.
    4-Press the “Up” or the “Down” button until you reach “Oil Expel”.
    5-Press the “Menu/Enter” button to select “Oil Expel” mode.

  • Do you repair other brands of autoclaves?

    Yes we offer onsite repair for a lot of different autoclave’s brand, Please check our autoclave page and for more information you can contact us